マッハ–ツェンダー干渉計(MZI)     


OL-MZI interferometer is used to scan OCT system. It contains a fixed arm difference MZI and a low noise photoelectric balance detector. The balance detector in the module is used to provide K clock signal. We can customize different wavelengths and different arm differences according to different needs of customers. The module is designed for shock absorption and heat insulation to ensure the stability of the module to the greatest extent. 



製品 モデル


ネーム モデル 価額
マッハ・ツェンダー干渉計   [PDF]  [RFQ]

OL-MZI-1300
Stock NO.: D80020017
[お問い合わせください]

パラメータ


特長

● Different wavelength can be customized

● Different arm difference MZI can be customized

● Output with Balanced Detector

● Design of shock absorption and heat insulation

● Compact structure

仕様

Product   Model

OL-MZI-1300

Wavelength

1225-1375nm

Free   Spectrum Range MZI Output

103.3GHz±5%

MZI arm difference

2mm (other arm differences can be   customized)

Optical Fiber Type

SMF-28PM

Optical Input

FC/APC

Detector Model

InGaAs   / PIN

Detector wavelength

800 - 1700nm

Balance detector bandwidth

200MHz

Saturated optical power

50mW@1300nm

Connector

SMA

Supply Voltage/Current

5V/0.5Amax

Shape Size

120*100*25mm

Support customization



Test Result

P19.jpg

Mach-Zehnder Interferometer Optical Path Map


応用 / 用途

● OCT

● Other Coherent fields